RecallHawk
Class II Recall

Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 1

Siemens Medical Solutions USA, Inc

Summary

The FDA issued a Class II for Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis z by Siemens Medical Solutions USA, Inc. Reason: Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently coo.

Details

Source

Device Recall

External ID

Z-1598-2024

Action Date

2024-04-24

Status

Ongoing

Category

device

Product Description

Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143

Lot/Code Info: Model/UDI-DI: Artis zee floor- 4056869010045 Artis zee ceiling- 4056869010052 Artis zee multi-purpose- 4056869010076 Artis zee biplane- 4056869010069 Artis zee floor MN- 4056869010090 Artis zee biplane MN- 4056869010106 Artis zeego- 4056869010083 Artis Q floor- 4056869009971 Artis Q ceiling- 4056869009988 Artis Q biplane- 4056869009995 Artis Q zeego- 4056869010007 Artis Q.zen floor- 4056869010014 Artis Q.zen ceiling- 4056869010021 Artis Q.zen biplane- 4056869010038

Quantity Affected: 583 units

Reason for Recall

Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-03-21

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 179 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Siemens Medical Solutions USA, Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Siemens Medical Solutions USA, Inc have FDA actions?

Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1598-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions