Class II Recall

DigitalDiagnost C50 1.1

Philips North America Llc

Summary

The FDA issued a Class II for DigitalDiagnost C50 1.1 by Philips North America Llc. Reason: Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect o.

Details

Source

Device Recall

External ID

Z-1563-2022

Action Date

2022-09-07

Status

Ongoing

Category

device

Product Description

DigitalDiagnost C50 1.1

Lot/Code Info: Model Number 712204 Serial Number 210182 210187 220002 210049 210051 210097 210052 210053 210055 220020

Quantity Affected: 11 systems (updated 1/3/2023)

Reason for Recall

Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Added 1/3/2023), Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.

Distribution

Nationwide Distribution including Puerto Rico

Type: FDA Mandated

Recall Initiated: 2022-02-25

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 164 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Philips North America Llc has 302 FDA actions in our database, including 302 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Philips North America Llc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Philips North America Llc have FDA actions?

Philips North America Llc has 302 FDA actions in our database, including 302 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1563-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions