RecallHawk
Class II Recall

Welch Allyn CONNEX Accessory Power Management Stand:

Baxter Healthcare Corporation

Summary

The FDA issued a Class II for Welch Allyn CONNEX Accessory Power Management Stand: by Baxter Healthcare Corporation. Reason: There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing th.

Details

Source

Device Recall

External ID

Z-1559-2025

Action Date

2025-04-23

Status

Ongoing

Category

device

Product Description

Welch Allyn CONNEX Accessory Power Management Stand:

Lot/Code Info: All Serial Numbers. Product Code: 7000-APM, UDI-DI: 00732094210613; Product Code: 75-HCA-CTB, UDI-DI: 00732094240597; Product Code: 75-HCA-MTB, UDI-DI: 00732094240603;

Quantity Affected: 114,202 units

Reason for Recall

There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

Distribution

Worldwide - US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-03-24

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 165 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Baxter Healthcare Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Baxter Healthcare Corporation have FDA actions?

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1559-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions