RecallHawk
Class II Recall

Merlin PCS 3650 programmer Model 3330 software

St. Jude Medical, Cardiac Rhythm Management Division

Summary

The FDA issued a Class II for Merlin PCS 3650 programmer Model 3330 software by St. Jude Medical, Cardiac Rhythm Management Division. Reason: There is a potential for programmer software (Merlin PCS) and remote monitoring software application (Merlin.net) to display overestimated predicted b.

Details

Source

Device Recall

External ID

Z-1495-2022

Action Date

2022-08-10

Status

Ongoing

Category

device

Product Description

Merlin PCS 3650 programmer Model 3330 software

Lot/Code Info: Software Model 3330 v26.0.1 rev 1 or earlier; UDI/DI 05414734509725

Quantity Affected: 29,996 programs

Reason for Recall

There is a potential for programmer software (Merlin PCS) and remote monitoring software application (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers.

Distribution

US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VI VT WA WI WV WY , Guam OUS: ALBANIA ALGERIA ANGOLA Argentina ARMENIA Australia AUSTRIA AZERBAIJAN BAHRAIN Bangladesh BELGIUM BENIN BOLIVIA BOSNIA-HERZ. BRAZIL BULGARIA Burnei Cambodia CAMEROON Canada Chile China Colombia COTE D'IVOIRE CUBA CYPRUS CZECH REPUBLIC DENMARK DOMINICAN REP. ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FAROE ISLANDS FINLAND FRANCE GEORGIA GERMANY GREECE GREENLAND GUATEMALA Hong Kong HUNGARY ICELAND India Indonesia IRELAND ISRAEL ITALY JORDAN KAZAKHSTAN KOSOVO KUWAIT KYRGYZSTAN LATVIA LEBANON LIBYA LIECHTENSTEIN LITHUANIA LUXEMBOURG MACEDONIA Malaysia MALTA MAURITIUS Mexico MONGOLIA MONTENEGRO MOROCCO Myanmar Nepal NETHERLANDS New Zealand NORWAY OMAN Pakistan PALESTINE PANAMA PARAGUAY PERU Philippines POLAND PORTUGAL QATAR ROMANIA RUSSIAN FED. SAUDI ARABIA SENEGAL Singapore SLOVAKIA SLOVENIA SOUTH AFRICA South Korea SPAIN Srilanka SUDAN SWEDEN SWITZERLAND Taiwan Thailand TRINIDAD TUNISIA TURKEY UKRAINE United Arab Emirates UNITED KINGDOM UZBEKISTAN VENEZUELA Vietnam VIRGIN ISLAND YEMEN ZAMBIA

Type: Voluntary: Firm initiated

Recall Initiated: 2022-06-17

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 97 device recalls issued in the same week, part of 403 device-related FDA actions this month.

St. Jude Medical, Cardiac Rhythm Management Division has 8 FDA actions in our database, including 8 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (St. Jude Medical, Cardiac Rhythm Management Division) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does St. Jude Medical, Cardiac Rhythm Management Division have FDA actions?

St. Jude Medical, Cardiac Rhythm Management Division has 8 FDA actions in our database, including 8 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1495-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions