RecallHawk
Class II Recall

Luminos dRF Max (VE10, VF10, VF11)

Siemens Medical Solutions USA, Inc

Summary

The FDA issued a Class II for Luminos dRF Max (VE10, VF10, VF11) by Siemens Medical Solutions USA, Inc. Reason: Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect .

Details

Source

Device Recall

External ID

Z-1480-2023

Action Date

2023-05-31

Status

Ongoing

Category

device

Product Description

Luminos dRF Max (VE10, VF10, VF11)

Lot/Code Info: Model: 10762471

Quantity Affected: 704 units in the United States (1945 units worldwide)

Reason for Recall

Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.

Distribution

US Nationwide - Worldwide Distribution

Type: FDA Mandated

Recall Initiated: 2022-11-23

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 141 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Siemens Medical Solutions USA, Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Siemens Medical Solutions USA, Inc have FDA actions?

Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1480-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions