Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE
Summary
The FDA issued a Class II for Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT by Medline Industries, LP. Reason: Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tea.
Details
Source
Device Recall
External ID
Z-1466-2026
Action Date
2026-03-04
Status
Ongoing
Category
device
Product Description
Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A
Lot/Code Info: SKU DYNDA2705 UDI-DI (ea) 10193489468632 UDI-DI (case) 40193489468633 Kit lots 24DBL787 24HBR630 25ABH699 25DBV209 25GBQ405 SKU DYNJ63347A UDI-DI (ea) 10193489278064 UDI-DI (case) 40193489278065 Kit lots 24AMC874 24AME139 24AMH820 24BMF097 24CMF691 24GMB769 24GMK036 24IMC124 24LMA110 25AMJ634 25GMD386 25IMB510 25JMI708
Quantity Affected: 1212
Reason for Recall
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Distribution
US Nationwide distribution.
Type: Voluntary: Firm initiated
Recall Initiated: 2026-01-08
Company
Northfield, IL
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 178 device recalls issued in the same week, part of 413 device-related FDA actions this month.
Medline Industries, LP has 442 FDA actions in our database, including 415 recalls and 27 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medline Industries, LP) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Medline Industries, LP have FDA actions?
Medline Industries, LP has 442 FDA actions in our database, including 415 recalls and 27 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-1466-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri