Summary

The FDA issued a Class II for Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Gu by Checkpoint Surgical Inc. Reason: Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed met.

Details

Source

Device Recall

External ID

Z-1466-2025

Action Date

2025-04-02

Status

Ongoing

Category

device

Product Description

Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Small Model/Catalog Number: REF# 9524 (Small) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.

Lot/Code Info: UDI-DI B13095240 Small Lot 4318 Expires 01Apr2026 Lot 4325 Expires 01Apr2026 Lot 4368 Expires 01Dec2026

Quantity Affected: 135

Reason for Recall

Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-02-24

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 216 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Checkpoint Surgical Inc has 2 FDA actions in our database, including 2 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Checkpoint Surgical Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Checkpoint Surgical Inc have FDA actions?

Checkpoint Surgical Inc has 2 FDA actions in our database, including 2 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1466-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions