RecallHawk
Class I Recall

Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010

Boston Scientific Corporation

Summary

The FDA issued a Class I for Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013 by Boston Scientific Corporation. Reason: An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a .

Details

Source

Device Recall

External ID

Z-1374-2024

Action Date

2024-04-03

Status

Ongoing

Category

device

Product Description

Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010

Lot/Code Info: a) M0013972001010, UDI/DI 00191506039332, ALL LOT CODES b) M0013972101010, UDI/DI 00191506043124, ALL LOT CODES

Quantity Affected: 985 units

Reason for Recall

An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.

Distribution

US

Type: Voluntary: Firm initiated

Recall Initiated: 2024-02-20

Company

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 163 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Boston Scientific Corporation has 399 FDA actions in our database, including 334 recalls and 65 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Boston Scientific Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Boston Scientific Corporation have FDA actions?

Boston Scientific Corporation has 399 FDA actions in our database, including 334 recalls and 65 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1374-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions