Class II Recall

VITROS Performance Verifier II

Ortho-Clinical Diagnostics

Summary

The FDA issued a Class II for VITROS Performance Verifier II by Ortho-Clinical Diagnostics. Reason: The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause err.

Details

Source

Device Recall

External ID

Z-1373-2024

Action Date

2024-04-03

Status

Ongoing

Category

device

Product Description

VITROS Performance Verifier II

Lot/Code Info: Product Code: 8231474; GTIN: 10758750004577; Lot No. R1176.

Quantity Affected: 11,224 units

Reason for Recall

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-02-27

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 163 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Ortho-Clinical Diagnostics has 14 FDA actions in our database, including 5 recalls and 9 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Ortho-Clinical Diagnostics) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Ortho-Clinical Diagnostics have FDA actions?

Ortho-Clinical Diagnostics has 14 FDA actions in our database, including 5 recalls and 9 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1373-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions