RecallHawk
Class II Recall

Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

Preat Corp

Summary

The FDA issued a Class II for Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967 by Preat Corp. Reason: Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement lengt.

Details

Source

Device Recall

External ID

Z-1321-2024

Action Date

2024-03-27

Status

Ongoing

Category

device

Product Description

Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

Lot/Code Info: Model Number: 9000967 UDI-DI code: 00842092165867 Lot Numbers: 235650, 234294, 233667, 233885, 233439, and 231489

Quantity Affected: 561 units

Reason for Recall

Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.

Distribution

U.S. Nationwide distribution in the states of AZ, OH and PA.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-02-09

Company

Preat Corp

Santa Maria, CA

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 158 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Preat Corp has 11 FDA actions in our database, including 11 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Preat Corp) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Preat Corp have FDA actions?

Preat Corp has 11 FDA actions in our database, including 11 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1321-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions