RecallHawk
Class II Recall

Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050,

ETAC A/S

Summary

The FDA issued a Class II for Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M by ETAC A/S. Reason: A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of.

Details

Source

Device Recall

External ID

Z-1309-2026

Action Date

2026-02-11

Status

Ongoing

Category

device

Product Description

Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7) M15002; Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength. Component: The deformation occurs between the bolt and the fork in the joint connecting the lifting bar and the arm on the Molift Mover.

Lot/Code Info: (1) Model/Catalog Number: M1501; UDI-DI: TBD; Serial Numbers: 601005, 601002, 611506, 621101, 621106, 621104, 621105, 621102, 611505, 611504, 611503, 611502, 611501, 601008, 611507, 601007, 601006, 601001, 621103, 601003, 601004; (2) Model/Catalog Number: M1502; UDI-DI: TBD; Serial Numbers: 611409, 611408, 611407, 611406, 611405, 611404, 611403, 611402, 611401, 611412, 611411, 611410; (3) Model/Catalog Number: M1504; UDI-DI: TBD; Serial Numbers: 600902, 600901, 611303, 611302, 611301; (4) Model/Catalog Number: M15050; UDI-DI: TBD; Serial Numbers: 586003, 586001, 585905, 585903, 585902, 585901, 598605, 586101, 585904, 603602, 591401, 586002, 586006, 586005, 586004, 598608, 598607, 585906, 591601, 598606, 598604, 598602, 598601, 596501, 603601, 598609, 598603; (5) Model/Catalog Number: M15151; UDI-DI: TBD; Serial Numbers: 599703, 599702, 599701; (6) Model/Catalog Number: M15001; UDI-DI: TBD; Serial Numbers: 584401; (7) Model/Catalog Number: M15002; UDI-DI: TBD; Serial Numbers: 591501, 591502;

Quantity Affected: 71

Reason for Recall

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Distribution

Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-11-17

Company

ETAC A/S

Gedved, N/A

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 207 device recalls issued in the same week, part of 403 device-related FDA actions this month.

ETAC A/S has 4 FDA actions in our database, including 4 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (ETAC A/S) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does ETAC A/S have FDA actions?

ETAC A/S has 4 FDA actions in our database, including 4 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1309-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions