RecallHawk
Class II Recall

BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitorin

CareFusion 303, Inc.

Summary

The FDA issued a Class II for BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX by CareFusion 303, Inc.. Reason: Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This.

Details

Source

Device Recall

External ID

Z-1231-2025

Action Date

2025-03-05

Status

Ongoing

Category

device

Product Description

BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration

Lot/Code Info: Catalog Number: 8100 UDI-DI codes: 10885403810046 10885403517723 Software Version: 12.1.2 Serial/Lot Numbers: Pending

Quantity Affected: 242,098 systems

Reason for Recall

Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.

Distribution

U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Guantanamo Bay, Cuba O.U.S.: Canada

Type: Voluntary: Firm initiated

Recall Initiated: 2025-01-23

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 156 device recalls issued in the same week, part of 413 device-related FDA actions this month.

CareFusion 303, Inc. has 105 FDA actions in our database, including 103 recalls and 2 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (CareFusion 303, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does CareFusion 303, Inc. have FDA actions?

CareFusion 303, Inc. has 105 FDA actions in our database, including 103 recalls and 2 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1231-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions