RecallHawk
Class II Recall

BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352;

CareFusion 303, Inc.

Summary

The FDA issued a Class II for BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation E by CareFusion 303, Inc.. Reason: Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or .

Details

Source

Device Recall

External ID

Z-1223-2024

Action Date

2024-03-13

Status

Ongoing

Category

device

Product Description

BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352;

Lot/Code Info: REF/UDI-DI: 323/10885403512667, 327/10885403477836, 352/10885403512674. All Using Software Version ES System Release 1.4.x, 1.5.x and 1.6.x

Quantity Affected: 20,443

Reason for Recall

Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or crash, which may result in delays in medication access, and potential data loss.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of VA, SC, IL, AL, KS, LA, GA, FL, CA, PR, PA, WV, WA, HI, NC, MD, AR, NY, ND, TX, NM, GU, MS, AZ, DC, AK, NE, DE, CO, SD, MO, NJ, OK, NV, OH, KY, WY, MT, UT, ID, MA, MI, MP, NH, IA, TN, CT, RI, IN, WI, OR, MN, VT, ME and the countries of BH, DE, GB, IT, JP, ES, GU, KR, BE, AR, BR, BS, CL, CO, MX, AE, AU, CA, EG, FR, HK, MY, NZ, QA, SA, TH, NL, MC, CH, FI, JO, KW, LB, OM.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-02-01

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 225 device recalls issued in the same week, part of 413 device-related FDA actions this month.

CareFusion 303, Inc. has 105 FDA actions in our database, including 103 recalls and 2 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (CareFusion 303, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does CareFusion 303, Inc. have FDA actions?

CareFusion 303, Inc. has 105 FDA actions in our database, including 103 recalls and 2 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1223-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions