Summary
The FDA issued a Class II for Colonoscope, Model Number CF-HQ190L. by Olympus Corporation of the Americas. Reason: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vib.
Details
Source
Device Recall
External ID
Z-1207-2024
Action Date
2024-03-06
Status
Ongoing
Category
device
Product Description
Colonoscope, Model Number CF-HQ190L.
Lot/Code Info: Model Number: CF-HQ190L; UDI DI: 04953170305115; Serial Number: 2082724, 2082896, 2082936, 2083461, 2083765, 2084560, 2085050, 2186261, 2187110, 2187274, 2187811, 2188037, 2188206, 2200530, 2201625, 2201677, 2201844, 2202345, 2288483, 2288940, 2289129, 2289391, 2289421, 2289572, 2290978, 2291472, 2303312, 2304024, 2304070, 2314598, 2315390, 2391750, 2391772, 2391857, 2392281, 2417328, 2417450, 2417702, 2417736, 2417755, 2418202, 2418754, 2418843, 2419006, 2419007, 2419105, 2510489, 2510795, 2510801, 2511164, 2512292, 2512458, 2512470, 2512569, 2512817, 2513285, 2524619, 2524916, 2625144, 2635605, 2635832, 2636020, 2636500, 2636782, 2637892, 2659060, 2659357, 2759754, 2761671, 2762243, 2772492, 2773247, 2773259, 2773324, 2773552, 2874236, 2875490, 2875615, 2875616, 2875656, 2876636, 2876811, 2876877, 2877237, 2877319, 2877901, 2978291, 2978331, 2978640, 2979287, 2979510, 2980281, 2980304, 2980541, 2981322, 2982038, 2982143;
Quantity Affected: 97
Reason for Recall
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Distribution
US Nationwide distribution.
Type: Voluntary: Firm initiated
Recall Initiated: 2024-01-11
Company
Center Valley, PA
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 205 device recalls issued in the same week, part of 413 device-related FDA actions this month.
Olympus Corporation of the Americas has 334 FDA actions in our database, including 334 recalls.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Olympus Corporation of the Americas) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Olympus Corporation of the Americas have FDA actions?
Olympus Corporation of the Americas has 334 FDA actions in our database, including 334 recalls and 0 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-1207-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;