RecallHawk
Class II Recall

EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.

Boston Scientific Corporation

Summary

The FDA issued a Class II for EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or by Boston Scientific Corporation. Reason: Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use.

Details

Source

Device Recall

External ID

Z-1153-2024

Action Date

2024-02-28

Status

Ongoing

Category

device

Product Description

EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.

Lot/Code Info: UPN M00509500, GTIN (UDI-DI) 08714729902713, Lot 31073152; UPN M00509510, GTIN (UDI-DI) 08714729902720, Lots 31221751, 31761882, 32409573.

Quantity Affected: 0 US; 74 OUS

Reason for Recall

Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may cause embedding with erosion into the gastric wall, tissue necrosis, infection, sepsis, and associate sequelae.

Distribution

International distribution to the country of Japan only.

Type: Voluntary: Firm initiated

Recall Initiated: 2023-12-19

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 140 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Boston Scientific Corporation has 399 FDA actions in our database, including 334 recalls and 65 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Boston Scientific Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Boston Scientific Corporation have FDA actions?

Boston Scientific Corporation has 399 FDA actions in our database, including 334 recalls and 65 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1153-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions