RecallHawk
Class I Recall

Cardiosave Rescue Model Nos. 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85

Datascope Corp.

Summary

The FDA issued a Class I for Cardiosave Rescue Model Nos. 0998-00-0800-75 0998-00-0800-83 0998-00-0800- by Datascope Corp.. Reason: An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may .

Details

Source

Device Recall

External ID

Z-1148-2023

Action Date

2023-03-15

Status

Ongoing

Category

device

Product Description

Cardiosave Rescue Model Nos. 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85

Lot/Code Info: Model UDI 0998-00-0800-75 10607567112312 0998-00-0800-83 10607567108407 0998-00-0800-85 10607567113449 All serial numbers

Quantity Affected: 4793 (US); 4407 (OUS)

Reason for Recall

An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.

Distribution

Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, and WV; and OUS (Foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BARBADOS, BELRUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire, CROATIA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE,GERMANY,GHANA,GUATEMALA,HONG KONG,HUNGARY,ICELAND,INDIA,INDONESIA IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO TUNISIA, TURKEY, TURKMENISTAN, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM UZBEKISTAN, VIETNAM, YEMEN AND ZAMBIA.

Type: Voluntary: Firm initiated

Recall Initiated: 2023-02-07

Company

Datascope Corp.

Mahwah, NJ

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 137 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Datascope Corp. has 58 FDA actions in our database, including 58 recalls.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Datascope Corp.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Datascope Corp. have FDA actions?

Datascope Corp. has 58 FDA actions in our database, including 58 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1148-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions