Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code
Summary
The FDA issued a Class II for Impella catheters - Intravascular micro axial blood pumps that support a patient by Abiomed, Inc.. Reason: New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, w.
Details
Source
Device Recall
External ID
Z-1074-2024
Action Date
2024-02-21
Status
Ongoing
Category
device
Product Description
Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 100080 and 0048-0045 - Impella CP with SmartAssist -- ***Added 2/20/24*** (10) 4413 and 004413 - Impella 2.5 Set (11) 5040 and 005040 - Impella 2.5 IMC Pump Set EU (12) 5060 and 005060 - Impella 5.0 IMC Pump Set EU (13) 5064 and 005064 - Impella 5.0 IMC Pump Set Canada (14) 1000115 - Impella CP Pump set, APAC (15) 1000211 - Impella 5.5 Smart Assist Set, JP (16) 1000302 - Impella CP with SmartAssist APAC (17) 1000361 - Impella 5.5 Set AU (18) 1000402 - Impella CP Smart Assist Set (19) 0046-0011 - Impella RP Pump Set, EU (20) 0046-0026 - Impella 5.0 Pump Set ROW (21) 0046-0037 - Impella 5.0 Pump Set APAC (22) 0046-0039 - Impella RP Set APAC (23) 004690-AU - Impella 5.0 Pump Set AU (24) 0048-0002 - Impella CP Pump Set, EU (25) 0048-0002-BR - Impella CP Pump Set BR (26) 0048-0004 - Impella CP Pump Set, Canada (27) 0048-0014 - Impella Smart Assist Set, EU (28) 0048-0024-JP - Impella CP Smart Assist Set, JP (29) 0048-0044 - Impella CP Smart Assist Set, Canada (30) 0048-0047- Impella CP Smart Assist Set APAC (31) 005048-JP - Impella 2.5 Pump Set, Japan (32) 005066-JP - Impella 5.0 Pump Set, Japan (33) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (34) 0550-0004 - Impella 5.5 with SmartAssist Set, CA
Lot/Code Info: UDI-DI (1) 00813502011081 (2) 00813502011180 (3) 00813502011227 (4) 00813502011029 (5) 00813502011869 (6) 00813502012811 (7) 00813502011531 and 00813502012828 (8) 00813502011388 (9) 00813502012279 and 00813502011876 All products are affected by IFU Update ***Added 2/20/24*** (10) 813502010947 (11) 4260113630136 (12) 4260113630174 (13) 813502011197 (14) 813502010473 (15) 813502012453 (16) 813502012767 (17) 813502012927 (18) 813502012873 (19) 4260113630273 (20) 813502011821 (21) 813502011937 (22) 813502011951 (23) N/A (24) 4260113630242 (25) 813502011777 (26) 813502011265 (27) 4260113630280 (28) 813502011609 (29) 813502012200 (30) 813502011944 (31) 813502010046 (32) 813502010053 (33) 813502011630 (34) 813502010466
Quantity Affected: 92515
Reason for Recall
New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR, IN, MX, TW, and VI.
Type: Voluntary: Firm initiated
Recall Initiated: 2023-12-27
Company
Danvers, MA
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 164 device recalls issued in the same week, part of 413 device-related FDA actions this month.
Abiomed, Inc. has 66 FDA actions in our database, including 61 recalls and 5 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Abiomed, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Abiomed, Inc. have FDA actions?
Abiomed, Inc. has 66 FDA actions in our database, including 61 recalls and 5 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-1074-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29