Class II Recall

ProxiDiagnost upgrade- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref:706150

Philips North America

Summary

The FDA issued a Class II for ProxiDiagnost upgrade- A Multi-functional general Radiography and Fluoroscopy (R by Philips North America. Reason: Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patients radiography (RAD) images when starting the next patient .

Details

Source

Device Recall

External ID

Z-1073-2024

Action Date

2024-02-21

Status

Ongoing

Category

device

Product Description

ProxiDiagnost upgrade- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref:706150

Lot/Code Info: UDI: N/A Serial Numbers US: 4 5 6 7 8 9 11 12 13 14 17 18 61 692222229 692222254 692222263 Serial Numbers OUS: 53 56 65 692222227

Quantity Affected: 20 units: 16 units US; 4 units OUS

Reason for Recall

Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patients radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)

Distribution

Nationwide Foreign: Australia Germany India Italy Malaysia Panama Thailand

Type: Voluntary: Firm initiated

Recall Initiated: 2023-12-15

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 164 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Philips North America has 126 FDA actions in our database, including 126 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Philips North America) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Philips North America have FDA actions?

Philips North America has 126 FDA actions in our database, including 126 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1073-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.