Allura Xper FD20; Model Number: 722012;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Summary

The FDA issued a Class II for Allura Xper FD20; Model Number: 722012; by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. Reason: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corr.

Details

Source

Device Recall

External ID

Z-1070-2026

Action Date

2026-01-21

Status

Ongoing

Category

device

Product Description

Allura Xper FD20; Model Number: 722012;

Lot/Code Info: Model Number: 722012; System Product Name: Allura Xper FD20; UDI-DI: 00884838059054; Serial Numbers: 2942, 1515, 1189, 1594, 535, 1964, 2156, 2742, 2092, 1673, 1803, 1655, 2020, 1973, 2037, 1853, 2098, 1248, 1082, 1576, 1057, 1129, 646, 930, 1421, 462, 203, 627, 2005, 1428, 353, 288, 191, 1994, 1802, 156, 853, 1283, 1956, 2361, 1980, 872, 931, 1295, 2513, 1090, 1374, 468, 1822, 812, 1379, 1817, 1580, 1050, 309, 144, 1841, 1845, 1468, 558, 1968, 1959, 2168, 1168, 1562, 967, 2114, 47, 167, 2091, 2053, 604, 603, 217, 2918, 373, 850, 1820, 1821, 1619;

Quantity Affected: 79 units

Reason for Recall

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-12-15

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 161 device recalls issued in the same week, part of 413 device-related FDA actions this month.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has 236 FDA actions in our database, including 203 recalls and 33 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. have FDA actions?

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has 236 FDA actions in our database, including 203 recalls and 33 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1070-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions