RecallHawk
Class II Recall

Discovery XR656HD. X-Ray imaging system.

GE Medical Systems, LLC

Summary

The FDA issued a Class II for Discovery XR656HD. X-Ray imaging system. by GE Medical Systems, LLC. Reason: The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to ackno.

Details

Source

Device Recall

External ID

Z-1044-2025

Action Date

2025-02-12

Status

Ongoing

Category

device

Product Description

Discovery XR656HD. X-Ray imaging system.

Lot/Code Info: GTINs: 00840682142175, 00840682142168

Quantity Affected: 3149 (1697 US; 1452 OUS)

Reason for Recall

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.

Distribution

Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.

Type: FDA Mandated

Recall Initiated: 2025-01-28

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 165 device recalls issued in the same week, part of 403 device-related FDA actions this month.

GE Medical Systems, LLC has 138 FDA actions in our database, including 121 recalls and 17 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (GE Medical Systems, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does GE Medical Systems, LLC have FDA actions?

GE Medical Systems, LLC has 138 FDA actions in our database, including 121 recalls and 17 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1044-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.