Class II Recall

Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit

Beckman Coulter, Inc.

Summary

The FDA issued a Class II for Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoas by Beckman Coulter, Inc.. Reason: Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a stand.

Details

Source

Device Recall

External ID

Z-1025-2026

Action Date

2026-01-14

Status

Ongoing

Category

device

Product Description

Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit

Lot/Code Info: A71460: UDI 15099590369224 A71461: UDI 15099590369248 A30260: UDI 15099590341602

Quantity Affected: 1272 units

Reason for Recall

Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.

Distribution

Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cabo Verde, Canada, Chile, China, Costa Rica, Croatia, Czechia, Denmark, Dominican Republic, Egypt, El Salvador, Eswatini, Ethiopia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Moldova, Republic of, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam, and Zimbabwe.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-11-26

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 98 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Beckman Coulter, Inc. has 158 FDA actions in our database, including 111 recalls and 47 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Beckman Coulter, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Beckman Coulter, Inc. have FDA actions?

Beckman Coulter, Inc. has 158 FDA actions in our database, including 111 recalls and 47 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1025-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions