RecallHawk
Class II Recall

Cardinal Health Jackson-Pratt Channel Drain, 32 Fr, REF JP-2292

Cardinal Health 200, LLC

Summary

The FDA issued a Class II for Cardinal Health Jackson-Pratt Channel Drain, 32 Fr, REF JP-2292 by Cardinal Health 200, LLC. Reason: Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including.

Details

Source

Device Recall

External ID

Z-0994-2023

Action Date

2023-01-25

Status

Ongoing

Category

device

Product Description

Cardinal Health Jackson-Pratt Channel Drain, 32 Fr, REF JP-2292

Lot/Code Info: UDI/DI 50885380182684(cs), 20885380182683(ea), Lot Numbers: N210543, N210460, N210468, N210469, N220050, N220099, N220139, N220163

Quantity Affected: 0

Reason for Recall

Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-12-14

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 149 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Cardinal Health 200, LLC has 153 FDA actions in our database, including 152 recalls and 1 clearance.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Cardinal Health 200, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Cardinal Health 200, LLC have FDA actions?

Cardinal Health 200, LLC has 153 FDA actions in our database, including 152 recalls and 1 device clearance.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0994-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.