RecallHawk
Class II Recall

PrisMax System, Product Code 955724

Baxter Healthcare Corporation

Summary

The FDA issued a Class II for PrisMax System, Product Code 955724 by Baxter Healthcare Corporation. Reason: This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value .

Details

Source

Device Recall

External ID

Z-0992-2022

Action Date

2022-05-04

Status

Ongoing

Category

device

Product Description

PrisMax System, Product Code 955724

Lot/Code Info: GTIN 07332414126018, All Serial Numbers

Quantity Affected: 296 units

Reason for Recall

This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.

Distribution

Worldwide distribution - US Nationwide and the countries of Mexico, Brazil, Jamaica, Trinidad and Tobago, Canada, Australia, Singapore, Hong Kong, Vietnam, New Zealand, Korea, Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-03-15

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 155 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Baxter Healthcare Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Baxter Healthcare Corporation have FDA actions?

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0992-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions