Artis icono floor in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single an

Siemens Medical Solutions USA, Inc

Summary

The FDA issued a Class II for Artis icono floor in combination with a Siemens Healthineers table or Trumpf/MA by Siemens Medical Solutions USA, Inc. Reason: Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table.

Details

Source

Device Recall

External ID

Z-0986-2022

Action Date

2022-04-27

Status

Ongoing

Category

device

Product Description

Artis icono floor in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327700

Lot/Code Info: S/N: Serial 170319 170026 170343 170367 170368 170305 170360 170307 170331 170349 170309 170310 170361 170351 170383 170023 170304 170328 170334 170336 170025 170346 170051 170358 170357 170333 170365 170332 170312 170027 170303 170378 170322 170053 170364 UDI: 04056869149325

Quantity Affected: 35 units

Reason for Recall

Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-02-25

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 150 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Siemens Medical Solutions USA, Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Siemens Medical Solutions USA, Inc have FDA actions?

Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0986-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions