RecallHawk
Class II Recall

CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US

Graphic Controls Acquisition Corporation

Summary

The FDA issued a Class II for CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US by Graphic Controls Acquisition Corporation. Reason: Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section.

Details

Source

Device Recall

External ID

Z-0935-2022

Action Date

2022-04-20

Status

Terminated

Category

device

Product Description

CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US

Lot/Code Info: All lots UDI: 03700506307357

Quantity Affected: 40 units

Reason for Recall

Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

Distribution

US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-03-15

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 137 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Graphic Controls Acquisition Corporation has 6 FDA actions in our database, including 6 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Graphic Controls Acquisition Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Graphic Controls Acquisition Corporation have FDA actions?

Graphic Controls Acquisition Corporation has 6 FDA actions in our database, including 6 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0935-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.