Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Articl
Summary
The FDA issued a Class II for Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Condu by B Braun Medical Inc. Reason: Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could.
Details
Source
Device Recall
External ID
Z-0873-2022
Action Date
2022-04-20
Status
Terminated
Category
device
Product Description
Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 04046964285608), 710200U (UDI 04046964285615), 710200K (UDI 04046963686338), 710500L (UDI Not applicable) and 710500K (UDI Not applicable).
Lot/Code Info: Lots 4/21 and 5/21
Quantity Affected: 95
Reason for Recall
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
Distribution
US Nationwide distribution.
Type: Voluntary: Firm initiated
Recall Initiated: 2022-01-28
Company
Bethlehem, PA
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 137 device recalls issued in the same week, part of 403 device-related FDA actions this month.
B Braun Medical Inc has 282 FDA actions in our database, including 269 recalls and 13 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (B Braun Medical Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does B Braun Medical Inc have FDA actions?
B Braun Medical Inc has 282 FDA actions in our database, including 269 recalls and 13 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-0873-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29