Summary
The FDA issued a Class II for IntelliVue Patient Monitor MX600. Product Number: 865242. by Philips North America Llc. Reason: Potential issue where the IntelliVue monitors did not alarm..
Details
Source
Device Recall
External ID
Z-0866-2026
Action Date
2025-12-10
Status
Ongoing
Category
device
Product Description
IntelliVue Patient Monitor MX600. Product Number: 865242.
Lot/Code Info: Product Number: 865242; UDI-DI: 00884838029095; All Serial No.
Quantity Affected: 1,913,441 units
Reason for Recall
Potential issue where the IntelliVue monitors did not alarm.
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Botswana, Brazil, Brunei, Bulgaria, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Equat Guinea, Estonia, Ethiopia, Faroe Islands, Finland, French Polynesia, France, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea Rep, Kuwait, Lao, Latvia, Lebanon, Lesotho, Liberia, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Ant, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, , R¿union, Romania, Russia, Saint Vincent, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, St. Pierre Mqln, Suriname, Sweden, Switzerland, Syrian Arab Rep, Taiwan, Tajikistan, Thailand, Trinidad Tobago, Tunisia, T¿rkiye, Turkmenistan, Uganda, Ukraine, United Kingdom, United States, Uruguay, Utd Arab Emir, Uzbekistan, Venezuela, Viet Nam, Virgin Isl (Br), Virgin Isl (US), White Russia, Yemen, Zambia, Zimbabwe.
Type: Voluntary: Firm initiated
Recall Initiated: 2025-10-31
Company
Cambridge, MA
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 336 device recalls issued in the same week, part of 403 device-related FDA actions this month.
Philips North America Llc has 302 FDA actions in our database, including 302 recalls.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Philips North America Llc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Philips North America Llc have FDA actions?
Philips North America Llc has 302 FDA actions in our database, including 302 recalls and 0 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-0866-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29