RecallHawk
Class I Recall

smiths medical Medfusion Model 3500 Syringe pump

Smiths Medical ASD Inc.

Summary

The FDA issued a Class I for smiths medical Medfusion Model 3500 Syringe pump by Smiths Medical ASD Inc.. Reason: Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm,.

Details

Source

Device Recall

External ID

Z-0863-2024

Action Date

2024-02-07

Status

Ongoing

Category

device

Product Description

smiths medical Medfusion Model 3500 Syringe pump

Lot/Code Info: All Medfusion Model 3500 pumps with software v3.X, v4.X, and v5.0.0 and was addressed in software v6.0.0 (2012) and all subsequent versions.

Quantity Affected: 93177 pumps

Reason for Recall

Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusion Restarted with Incorrect Parameters, 3. Screen Lock, 4. Interruption of Bolus or Loading Dose Delivery, 5. Pump Displays Incorrect Bolus/Loading Dose, 6. Loading/Bolus Dose Below the Minimum Recommended Rate, 7. Motor Rate Error, 8. Incorrect Recall Last Settings, 9. Corrupt Configuration, 10. Auto Lock, 11. Toolbox Configuration Loading Dose Time Values. Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software (v6.0.0) installed on your pumps.

Distribution

worldwide

Type: Voluntary: Firm initiated

Recall Initiated: 2023-12-19

Company

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 221 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Smiths Medical ASD Inc. has 164 FDA actions in our database, including 162 recalls and 2 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Smiths Medical ASD Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Smiths Medical ASD Inc. have FDA actions?

Smiths Medical ASD Inc. has 164 FDA actions in our database, including 162 recalls and 2 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0863-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions