RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.
Summary
The FDA issued a Class II for RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4 by RAYSEARCH LABORATORIES AB. Reason: An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Sup.
Details
Source
Device Recall
External ID
Z-0849-2022
Action Date
2022-04-06
Status
Ongoing
Category
device
Product Description
RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35, 5.0.3.17, 6.0.0.24, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 11.0.0.951 or 11.0.1.29
Lot/Code Info: Product name (build number) UDI-DI RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35) N/A RayStation 5 Service Pack 3 (5.0.3.17) 07350002010020 RayStation 6/RayPlan 2 (6.0.0.24) 07350002010013 RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2) 07350002010082 RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7) 07350002010075 RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6) 07350002010242 RayStation/RayPlan 7 (7.0.0.19) 07350002010068 RayStation/RayPlan 8A (8.0.0.61) 07350002010112 RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) 07350002010136 RayStation/RayPlan 8B (8.1.0.47) 07350002010129 RayStation/RayPlan 8B Service Pack 1 (8.1.1.8) 07350002010204 RayStation/RayPlan 8B Service Pack 2 (8.1.2.5) 07350002010235 RayStation/RayPlan 9A (9.0.0.113) 07350002010174 RayStation/RayPlan 9B (9.1.0.933) 07350002010266 RayStation/RayPlan 9B Service Pack 1 (9.2.0.483) 07350002010297 RayStation/RayPlan 10A (10.0.0.1154) 07350002010303 RayStation/RayPlan 10A Service Pack 1 (10.0.1.52) 07350002010365 RayStation/RayPlan 10B (10.1.0.613) 07350002010310 RayStation 11A (11.0.0.951) 07350002010389 RayStation 11A Service Pack 1 (11.0.1.29) 07350002010433
Quantity Affected: 478 licenses US
Reason for Recall
An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in the voxel can be both under- and overestimated
Distribution
US Nationwide.
Type: Voluntary: Firm initiated
Recall Initiated: 2021-09-22
Company
Stockholm, N/A
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 123 device recalls issued in the same week, part of 403 device-related FDA actions this month.
RAYSEARCH LABORATORIES AB has 51 FDA actions in our database, including 51 recalls.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (RAYSEARCH LABORATORIES AB) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does RAYSEARCH LABORATORIES AB have FDA actions?
RAYSEARCH LABORATORIES AB has 51 FDA actions in our database, including 51 recalls and 0 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-0849-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29