RecallHawk
Class II Recall

PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number

Becton Dickinson & Company

Summary

The FDA issued a Class II for PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe by Becton Dickinson & Company. Reason: An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush .

Details

Source

Device Recall

External ID

Z-0736-2024

Action Date

2024-01-24

Status

Ongoing

Category

device

Product Description

PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547

Lot/Code Info: UDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304

Quantity Affected: 5,085,120 units

Reason for Recall

An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.

Distribution

US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV.

Type: Voluntary: Firm initiated

Recall Initiated: 2023-11-16

Company

Becton Dickinson & Company

Franklin Lakes, NJ

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 203 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Becton Dickinson & Company has 39 FDA actions in our database, including 39 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Becton Dickinson & Company) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Becton Dickinson & Company have FDA actions?

Becton Dickinson & Company has 39 FDA actions in our database, including 39 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0736-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.