2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare

Fresenius Medical Care Holdings, Inc.

Summary

The FDA issued a Class II for 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and by Fresenius Medical Care Holdings, Inc.. Reason: (1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive t.

Details

Source

Device Recall

External ID

Z-0734-2023

Action Date

2022-12-28

Status

Ongoing

Category

device

Product Description

2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR

Lot/Code Info: Software issue in SW versions 2.72 and earlier with CDX. UDI: (1) 00840861100897; (2) 00840861100910; (3) 00840861102082; (4) 00840861102099

Quantity Affected: 2008T with CDX= 70,552 2008T Bluestar with CDX= 25,205

Reason for Recall

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-07-27

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 404 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Fresenius Medical Care Holdings, Inc. has 69 FDA actions in our database, including 69 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Fresenius Medical Care Holdings, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Fresenius Medical Care Holdings, Inc. have FDA actions?

Fresenius Medical Care Holdings, Inc. has 69 FDA actions in our database, including 69 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0734-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions