Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Summary

The FDA issued a Class II for Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. Reason: The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to .

Details

Source

Device Recall

External ID

Z-0719-2026

Action Date

2025-12-03

Status

Ongoing

Category

device

Product Description

Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;

Lot/Code Info: Model Number: 722234;

Quantity Affected: 172 units

Reason for Recall

The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.

Distribution

Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY, OH, OK, PA, RI, TX, UT, VA, WA, WI, WV, WY; International: Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Netherlands, Norway, Philippines, Poland, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Vietnam;

Type: Voluntary: Firm initiated

Recall Initiated: 2025-10-02

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 254 device recalls issued in the same week, part of 413 device-related FDA actions this month.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has 236 FDA actions in our database, including 203 recalls and 33 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. have FDA actions?

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has 236 FDA actions in our database, including 203 recalls and 33 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0719-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions