RecallHawk
Class II Recall

BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitat

Becton Dickinson & Co.

Summary

The FDA issued a Class II for BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin by Becton Dickinson & Co.. Reason: BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 m.

Details

Source

Device Recall

External ID

Z-0621-2025

Action Date

2024-12-11

Status

Ongoing

Category

device

Product Description

BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263

Lot/Code Info: UDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025).

Quantity Affected: 13,190 units

Reason for Recall

BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility

Distribution

Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-11-06

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 225 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Becton Dickinson & Co. has 73 FDA actions in our database, including 73 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Becton Dickinson & Co.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Becton Dickinson & Co. have FDA actions?

Becton Dickinson & Co. has 73 FDA actions in our database, including 73 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0621-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.