RecallHawk
Class II Recall

CareLink Clinic, REF: MMT-7350

Medtronic MiniMed, Inc.

Summary

The FDA issued a Class II for CareLink Clinic, REF: MMT-7350 by Medtronic MiniMed, Inc.. Reason: Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on.

Details

Source

Device Recall

External ID

Z-0594-2026

Action Date

2025-12-03

Status

Ongoing

Category

device

Product Description

CareLink Clinic, REF: MMT-7350

Lot/Code Info: UDI-DI: 0763000B00008748K. Software Versions: 4.2B Software Available: US: Between September 4th, 2025, and October 3rd, 2025. OUS: Between September 11th, 2025, and October 4th, 2025. US (Clinical): Between August 21st, 2025, and October 3rd, 2025. OUS (Clinical): Between August 21st, 2025, and October 4th, 2025.

Quantity Affected: 12126

Reason for Recall

Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview graph may lead to hypoglycemia or hyperglycemia.

Distribution

Worldwide Distribution: US (nationwide) including states of: TX, NC, MN, CA, VA, ID, WA, GA, MI, MA, OH, IL, IA, CT, FL, PA, NY, TN, AZ, LA, AL, NJ, CO, SC, MS, OR, MT, MD, KS, AK, NH, NE, WY, ND, IN, MO, WI, KY, UT, NV, SD, WV, NM, HI, AR, OK, VT, RI; and OUS (foreign) to countries of: Australia (AU), Sweden (SE), Belgium (BE), Great Britain (GB), Netherlands (NL), Finland (FI), Canada (CA), Germany (DE), Spain (ES), Switzerland (CH), Italy (IT), South Africa (ZA), South Korea (KR), Singapore (SG), New Zealand (NZ), Poland (PL), Israel (IL), Denmark (DK), Norway (NO), Luxembourg (LU), Lithuania (LT), Czech Republic ( CZ), Mexico (MX), Brazil (BR), Colombia (CO), Slovakia (SK), Ireland (IE), Argentina (AR), Greece (GR), Kuwait (KW), Portugal (PT), Iceland (IS), Japan (JP), Turkey (TR), Austria (AT), Qatar (QA), Hong Kong (HK), Chile (CL), Ukraine (UA), Hungary (HU), Estonia (EE), RS, Latvia (LV), Taiwan (TW), Saudia Arabia (SA), Bahrain (BH), Oman (OM), Romania (RO), India (IN), Uruguay (UY), Malaysia (MY), United Arab Emirates (AE), Algeria (DZ), Croatia (HR), France (FR), Russia (RU), Cyprus (CY), Thailand (TH), Egypt (EG), Bosnia and Herzegovina (BA), Libya (LY), and Iraq (IQ)

Type: Voluntary: Firm initiated

Recall Initiated: 2025-10-21

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 254 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Medtronic MiniMed, Inc. has 43 FDA actions in our database, including 37 recalls and 6 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medtronic MiniMed, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medtronic MiniMed, Inc. have FDA actions?

Medtronic MiniMed, Inc. has 43 FDA actions in our database, including 37 recalls and 6 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0594-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions