RecallHawk
Class II Recall

Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product

Olympus Corporation of the Americas

Summary

The FDA issued a Class II for Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V M by Olympus Corporation of the Americas. Reason: Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF du.

Details

Source

Device Recall

External ID

Z-0579-2026

Action Date

2025-12-03

Status

Ongoing

Category

device

Product Description

Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Lot/Code Info: Model/Catalog Number: TJF-Q190V; UDI: 04953170405563 / 04953170452024; Serial Numbers: All;

Quantity Affected: 7023 units

Reason for Recall

Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-10-14

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 254 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Olympus Corporation of the Americas has 334 FDA actions in our database, including 334 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Olympus Corporation of the Americas) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Olympus Corporation of the Americas have FDA actions?

Olympus Corporation of the Americas has 334 FDA actions in our database, including 334 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0579-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions