RecallHawk
Class II Recall

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4482EJ; use in Peritoneal Dialysis

Baxter Healthcare Corporation

Summary

The FDA issued a Class II for Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C44 by Baxter Healthcare Corporation. Reason: There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set..

Details

Source

Device Recall

External ID

Z-0560-2025

Action Date

2024-12-04

Status

Ongoing

Category

device

Product Description

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4482EJ; use in Peritoneal Dialysis

Lot/Code Info: UDI/DI 00085412489643, Lot/Serial Numbers: H24C19063, H24G22092, H24G25061

Quantity Affected: 22392 units

Reason for Recall

There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.

Distribution

Worldwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-10-23

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 162 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Baxter Healthcare Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Baxter Healthcare Corporation have FDA actions?

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0560-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions