RecallHawk
Class I Recall

OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pum

Abiomed, Inc.

Summary

The FDA issued a Class I for OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. by Abiomed, Inc.. Reason: The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a pat.

Details

Source

Device Recall

External ID

Z-0550-2022

Action Date

2022-02-09

Status

Terminated

Category

device

Product Description

OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.

Lot/Code Info: UDI 00860001797909 Software Version Number V-1.3.15.0, V-1.4.6.0 and V.1.4.8.0 Serial Numbers 20200231-04 20200249-01 20200249-05 20200309-02 20200339-01 20200339-03 20200352-01 20200442-01 20200442-02 20200442-05 20210768-01 20210795-02 20210870-01 20210880-01 20210890-01 20210894-01 20210895-01 20210896-01 20210897-01 20210795-04

Quantity Affected: 42 Units

Reason for Recall

The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.

Distribution

US distribution to FL, GA, IN, MD, NC, PA, TX.

Type: Voluntary: Firm initiated

Recall Initiated: 2021-12-14

Company

Abiomed, Inc.

Danvers, MA

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 95 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Abiomed, Inc. has 66 FDA actions in our database, including 61 recalls and 5 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Abiomed, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Abiomed, Inc. have FDA actions?

Abiomed, Inc. has 66 FDA actions in our database, including 61 recalls and 5 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0550-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions