RecallHawk
Class II Recall

CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and

Immuno-Mycologics, Inc

Summary

The FDA issued a Class II for CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detect by Immuno-Mycologics, Inc. Reason: The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 9.

Details

Source

Device Recall

External ID

Z-0545-2022

Action Date

2022-02-09

Status

Terminated

Category

device

Product Description

CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)00816387020919

Lot/Code Info: F1021035 and F1021036

Quantity Affected: 411 kits

Reason for Recall

The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.

Distribution

US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-01-03

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 95 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Immuno-Mycologics, Inc has 3 FDA actions in our database, including 3 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Immuno-Mycologics, Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Immuno-Mycologics, Inc have FDA actions?

Immuno-Mycologics, Inc has 3 FDA actions in our database, including 3 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0545-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.