Summary

The FDA issued a Class I for Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ by Medline Industries, LP. Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypo.

Details

Source

Device Recall

External ID

Z-0451-2026

Action Date

2025-11-19

Status

Ongoing

Category

device

Product Description

Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY

Lot/Code Info: DR SALMA'S SPINE DYNJ909111F UDI-DI 40198459130794 (case), 10198459130793 (ea) Lots 25FBE036 25GBV759; ORTHO SPINE FUMICH DYNJ901047R UDI-DI 40198459130466 (case), 10198459130465 (ea) Lots 25FBR230 25GBU693; PATTERSON TOTAL JOINT DYNJ909117G UDI-DI 40198459130480 (case), 10198459130489 (ea) Lots 25FBO307 25GBN625; TOTAL JOINT DYNJ902308K UDI-DI 40198459130596 (case), 10198459130595 (ea) Lots 25FBB321 25GBU688 KYPHOPLASTY DYNJ909113F UDI-DI 40198459130459 (case), 10198459130458 (ea) Lot 25IBM788

Quantity Affected: 96,406 total units

Reason for Recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Distribution

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Type: Voluntary: Firm initiated

Recall Initiated: 2025-10-10

Company

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 148 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Medline Industries, LP has 442 FDA actions in our database, including 415 recalls and 27 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medline Industries, LP) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medline Industries, LP have FDA actions?

Medline Industries, LP has 442 FDA actions in our database, including 415 recalls and 27 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0451-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions