RecallHawk
Class I Recall

BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100,

CareFusion 303, Inc.

Summary

The FDA issued a Class I for BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 4 by CareFusion 303, Inc.. Reason: If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregula.

Details

Source

Device Recall

External ID

Z-0430-2026

Action Date

2025-11-26

Status

Ongoing

Category

device

Product Description

BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide

Lot/Code Info: All Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743

Quantity Affected: 2,574,396

Reason for Recall

If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-10-17

Company

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 222 device recalls issued in the same week, part of 413 device-related FDA actions this month.

CareFusion 303, Inc. has 105 FDA actions in our database, including 103 recalls and 2 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (CareFusion 303, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does CareFusion 303, Inc. have FDA actions?

CareFusion 303, Inc. has 105 FDA actions in our database, including 103 recalls and 2 device clearances.