RecallHawk
Class II Recall

TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev

Integra LifeSciences Corp. (NeuroSciences)

Summary

The FDA issued a Class II for TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product N by Integra LifeSciences Corp. (NeuroSciences). Reason: It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the .

Details

Source

Device Recall

External ID

Z-0420-2026

Action Date

2025-11-12

Status

Ongoing

Category

device

Product Description

TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;

Lot/Code Info: Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; UDI-DI: 10846835021103;

Quantity Affected: 1198 units

Reason for Recall

It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-10-03

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 185 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Integra LifeSciences Corp. (NeuroSciences) has 30 FDA actions in our database, including 30 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Integra LifeSciences Corp. (NeuroSciences)) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Integra LifeSciences Corp. (NeuroSciences) have FDA actions?

Integra LifeSciences Corp. (NeuroSciences) has 30 FDA actions in our database, including 30 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0420-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions