Medline convenience kits containing BD Urinary Catheter Trays that include CliniClean Pericare Wipes DAVINCI DYNJ9
Summary
The FDA issued a Class II for Medline convenience kits containing BD Urinary Catheter Trays that include Clini by Medline Industries, LP. Reason: Convenience kits contain additional product codes and lots of recalled CliniClean Pericare Wipes identified by manufacturer. Wipes may be contaminated.
Details
Source
Device Recall
External ID
Z-0418-2027
Action Date
2026-11-18
Status
Ongoing
Category
device
Product Description
Medline convenience kits containing BD Urinary Catheter Trays that include CliniClean Pericare Wipes DAVINCI DYNJ906763C GYN ABDOMINAL ONCOLOGY DYNJ911414A GYN LAPAROSCOPIC PACK DYNJ909394B GYN LAPAROSCOPY DYNJ903449D DYNJ904701J GYN ROBOTIC CDS984076J DYNJ910181A DYNJ910181C GYN ROBOTIC ADD ON PACK DYNJ43837D OH GYN/LAP ROBOT PACK DYNJ89759 ROBOT ME RF DYNJ900203N ROBOTIC DYNJ903485K ROBOTIC COLON CDS CDS984953J ROBOTIC GU/GYN CDS983393L CDS983393M ROBOTIC GYN DYNJ905230K SOLK GYN-LAP ROBOT PACK DYNJT10414 W&I ROBOT DYNJ910664A WPH GYN LAPAROSCOPY PACK DYNJ89781 DYNJ89781A
Lot/Code Info: DYNJ906763C UDI-DI 10198459551833 Lots 25LBR121 26CBS290 DYNJ911414A UDI-DI 10198459670428 Lot 26FMI519 DYNJ909394B UDI-DI 10198459556579 Lots 25KBE469 26ABO368 DYNJ903449D UDI-DI 10195327407520 Lot 26CBN427 DYNJ904701J UDI-DI 10198459373060 Lot 26BDA186 CDS984076J UDI-DI 10198459283147 Lots 26CBD748 26DBA124 26DBC215 DYNJ910181A UDI-DI 10198459556586 Lots 25KLA664 26ALA732 26ALA897 DYNJ910181C UDI-DI 10198459780950 Lot 26ELA453 DYNJ43837D UDI-DI 10195327657222 Lot 26EMJ002 DYNJ89759 UDI-DI 10198459289286 Lots 25KBK074 25KBQ997 26ABN671 DYNJ900203N UDI-DI 10198459510144 Lots 26ABX032 26BBC420 26CBS914 DYNJ903485K UDI-DI 10198459255083 Lot 26BDA988 CDS984953J UDI-DI 10198459641367 Lots 26ABP040 26DBB664 CDS983393L UDI-DI 10198459224614 Lots 25LBC619 25LBH230 26ABI772 CDS983393M UDI-DI 10198459641350 Lot 26DBA175 DYNJ905230K UDI-DI 10198459520532 Lots 26CBF509 26CBQ707 DYNJT10414 UDI-DI 10198459792441 Lot 26FBE035 DYNJ910664A UDI-DI 10198459178672 Lot 26ABB917 DYNJ89781 UDI-DI 10198459289521 Lots 25LBR578 26CBK466 DYNJ89781A UDI-DI 10198459794933 Lot 26FBL798
Quantity Affected: 7,043 kits
Reason for Recall
Convenience kits contain additional product codes and lots of recalled CliniClean Pericare Wipes identified by manufacturer. Wipes may be contaminated with a lubricant used in the manufacturing process. Use of affected wipes may cause localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare may increase the risk of localized infection or catheter-associated urinary tract infection.
Distribution
US Nationwide distribution in the states of AL, AR, CA, FL, GA, IL, IN, LA, MA, MN, MO, MS, MT, NC, NE, NV, NY, OH, OK, RI, SC, TN, TX, WI.
Type: Voluntary: Firm initiated
Recall Initiated: 2026-09-24
Company
Northfield, IL
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 188 device recalls issued in the same week, part of 475 device-related FDA actions this month.
Medline Industries, LP has 630 FDA actions in our database, including 599 recalls and 31 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medline Industries, LP) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Medline Industries, LP have FDA actions?
Medline Industries, LP has 630 FDA actions in our database, including 599 recalls and 31 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-0418-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29