RecallHawk
Class II Recall

Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 973

Medtronic Navigation, Inc.

Summary

The FDA issued a Class II for Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth A by Medtronic Navigation, Inc.. Reason: A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instru.

Details

Source

Device Recall

External ID

Z-0417-2022

Action Date

2021-12-29

Status

Ongoing

Category

device

Product Description

Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundle Autoguide Reg Kit.

Lot/Code Info: All lot numbers

Quantity Affected: 98

Reason for Recall

A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy is experienced, it may result in prolonged procedure and tissue injury. Per 11/24/21 firm email: The Bundle Autoguide Reg Kit does not have a model number. The Bundle Autoguide Reg Kit is internal terminology and not visible to the customer and are used as sales and shipping configuration only. When this sales bundle is chosen by an internal Medtronic Sales Representative the devices included in the bundle are individually packaged and labeled and sent to the customer in a single shipment. The devices included in the bundle are ordered, packed, and shipped as individual medical devices with unique model numbers, which is the reason that the Autoguide Tracker (28248) and the Stealth Autoguide Basic Instrument Kit (9736188) are the only part numbers subject to this FCA.

Distribution

US: FL, NE, CA, MN, TX, OH, IL, AZ, KS, KY, OR, OK, WA, NV, MO, IN, MI, UT, MS, VA, SC. OUS: Canada, Sweden, Italy, France, Germany, Spain, Czech Republic, United Kingdom, India, Nepal, Japan, Turkey, China

Type: Voluntary: Firm initiated

Recall Initiated: 2021-11-09

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 113 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Medtronic Navigation, Inc. has 23 FDA actions in our database, including 13 recalls and 10 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medtronic Navigation, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medtronic Navigation, Inc. have FDA actions?

Medtronic Navigation, Inc. has 23 FDA actions in our database, including 13 recalls and 10 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0417-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.