RecallHawk
Class II Recall

Medline convenience kits containing BD Urinary Catheter Trays that include CliniClean Pericare Wipes CHEST/BREAST

Medline Industries, LP

Summary

The FDA issued a Class II for Medline convenience kits containing BD Urinary Catheter Trays that include Clini by Medline Industries, LP. Reason: Convenience kits contain additional product codes and lots of recalled CliniClean Pericare Wipes identified by manufacturer. Wipes may be contaminated.

Details

Source

Device Recall

External ID

Z-0415-2027

Action Date

2026-11-18

Status

Ongoing

Category

device

Product Description

Medline convenience kits containing BD Urinary Catheter Trays that include CliniClean Pericare Wipes CHEST/BREAST DYNJ908433F COLON DYNJ904031C DIAGNOSTIC LAP DYNJ900573I EAST MAJOR DYNJ900958C GENERAL LAPAROTOMY PACK- ESCO PHS825461003F GENERAL LAPAROTOMY POWAY DYNJT3500A GYN LAPAROSCOPY CDS CDS985204B LAP COLON DYNJ901360G MAJOR PROCEDURE CDS982040U OPERATIVE LAP P&S SURGICAL DYNJ89608

Lot/Code Info: DYNJ908433F UDI-DI 10198459544293 Lots 26AMF626 26AMG869 26DMD786 DYNJ904031C UDI-DI 10198459223518 Lot 26CBN236 DYNJ900573I UDI-DI 10198459551840 Lot 26CBR773 DYNJ900958C UDI-DI 10195327599454 Lots 25LBR186 26CBR774 PHS825461003F UDI-DI 10198459381249 Lots 26DMH776 26FME340 DYNJT3500A UDI-DI 10198459380907 Lot 26CMJ848 CDS985204B UDI-DI 10889942942039 Lot 25LBE036 DYNJ901360G UDI-DI 10198459224836 Lot 26CBM726 CDS982040U UDI-DI 10198459385421 Lot 25LMH746 DYNJ89608 UDI-DI 10198459279324 Lots 25KBG342 25KBO745 25LBP821

Quantity Affected: 7,043 kits

Reason for Recall

Convenience kits contain additional product codes and lots of recalled CliniClean Pericare Wipes identified by manufacturer. Wipes may be contaminated with a lubricant used in the manufacturing process. Use of affected wipes may cause localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare may increase the risk of localized infection or catheter-associated urinary tract infection.

Distribution

US Nationwide distribution in the states of AL, AR, CA, FL, GA, IL, IN, LA, MA, MN, MO, MS, MT, NC, NE, NV, NY, OH, OK, RI, SC, TN, TX, WI.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-09-24

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 188 device recalls issued in the same week, part of 475 device-related FDA actions this month.

Medline Industries, LP has 630 FDA actions in our database, including 599 recalls and 31 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medline Industries, LP) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medline Industries, LP have FDA actions?

Medline Industries, LP has 630 FDA actions in our database, including 599 recalls and 31 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0415-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.