RecallHawk
Class II Recall

Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China

Vantive US Healthcare LLC

Summary

The FDA issued a Class II for Sharesource Connectivity Platform for Use with Homechoice Claria Product Code by Vantive US Healthcare LLC. Reason: Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to .

Details

Source

Device Recall

External ID

Z-0409-2026

Action Date

2025-11-05

Status

Ongoing

Category

device

Product Description

Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)

Lot/Code Info: UDI-DI 15413765545723 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)

Quantity Affected: 1

Reason for Recall

Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician's remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Hong Kong, Korea, New Zealand, Taiwan, Thailand, France, Germany, Ireland, Netherlands, Italy, Spain, Turkey, United Kingdom, Chile, and Colombia.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-08-28

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 284 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Vantive US Healthcare LLC has 13 FDA actions in our database, including 12 recalls and 1 clearance.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Vantive US Healthcare LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Vantive US Healthcare LLC have FDA actions?

Vantive US Healthcare LLC has 13 FDA actions in our database, including 12 recalls and 1 device clearance.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0409-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions