RecallHawk
Class II Recall

OCS Liver System for organ transport OCS Solution Delivery Cassettes Model/Catalog Number: 1467 Product Descript

TransMedics, Inc.

Summary

The FDA issued a Class II for OCS Liver System for organ transport OCS Solution Delivery Cassettes Model/C by TransMedics, Inc.. Reason: Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant.

Details

Source

Device Recall

External ID

Z-0391-2027

Action Date

2026-11-18

Status

Ongoing

Category

device

Product Description

OCS Liver System for organ transport OCS Solution Delivery Cassettes Model/Catalog Number: 1467 Product Description: OCS Solution Delivery Cassettes The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. Each OCS Solution Delivery Cassettes contains three SDS cassettes. The BD syringe is a component of the SDS Cassette PN 100004459

Lot/Code Info: Lot Code: Lot: 4460-4-737189; Expiration Date: 2028-10-08; UDI: +B00944605/$$74460-4-737189/14D20281008/Q3L; Lot: 4460-4-738008; Expiration Date: 2028-11-07; UDI: +B00944605/$$74460-4-738008/14D20281107/Q3C; Lot: 4460-4-738009; Expiration Date: 2028-11-07; UDI: +B00944605/$$74460-4-738009/14D20281107/Q3D; Lot: 4460-4-739728; Expiration Date: 2029-01-14; UDI: +B00944605/$$74460-4-739728/14D20290114/Q3K;

Quantity Affected: 55 OCS Solution Delivery Cassettes

Reason for Recall

Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks.

Distribution

Worldwide - US Nationwide distribution in the state of NE and the country of Italy.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-11-20

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 188 device recalls issued in the same week, part of 475 device-related FDA actions this month.

TransMedics, Inc. has 12 FDA actions in our database, including 10 recalls and 2 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (TransMedics, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does TransMedics, Inc. have FDA actions?

TransMedics, Inc. has 12 FDA actions in our database, including 10 recalls and 2 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0391-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.