RecallHawk
Class II Recall

Sheehy VENTILATION TUBE 5PK BUTTON 1.27 FLPL, REF: 1013301, Lot# 0231048279 (GTIN 20763000281636), Lot# 0231048280 (GTIN

Medtronic Xomed, Inc.

Summary

The FDA issued a Class II for Sheehy VENTILATION TUBE 5PK BUTTON 1.27 FLPL, REF: 1013301, Lot# 0231048279 (GTI by Medtronic Xomed, Inc.. Reason: Ventilation tubes, that provide a passageway for movement of air between the auditory canal and the middle ear, were not subjected to the required ste.

Details

Source

Device Recall

External ID

Z-0387-2027

Action Date

2026-11-18

Status

Ongoing

Category

device

Product Description

Sheehy VENTILATION TUBE 5PK BUTTON 1.27 FLPL, REF: 1013301, Lot# 0231048279 (GTIN 20763000281636), Lot# 0231048280 (GTIN 20763000281636).

Lot/Code Info: UDI-DI: 20763000281636. Lots: 0231048279, 0231048280

Quantity Affected: 4000

Reason for Recall

Ventilation tubes, that provide a passageway for movement of air between the auditory canal and the middle ear, were not subjected to the required sterilization process, which after implantation could cause infection and subsequent inflammation potentially leading to temporary hearing impairment, otorrhea (fluid discharge) and the need for antibiotics.

Distribution

Worldwide - US Nationwide distribution including in the states of NY, AL, VA, MI, AZ, WA, TX, MA, IL, ME, NM, CA, IN, NC, OH, MO, MN, NH, IA, NV, GA, PA, FL, TN, WV, SC, OK, NE, OR, SD, AK, LA, CO, WI, MD, KS, UT, ID, WY, NJ and the countries of Canada, Australia, New Zealand, Hong Kong.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-09-24

Company

Medtronic Xomed, Inc.

Jacksonville, FL

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 188 device recalls issued in the same week, part of 475 device-related FDA actions this month.

Medtronic Xomed, Inc. has 18 FDA actions in our database, including 14 recalls and 4 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medtronic Xomed, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medtronic Xomed, Inc. have FDA actions?

Medtronic Xomed, Inc. has 18 FDA actions in our database, including 14 recalls and 4 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0387-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions