RecallHawk
Class II Recall

Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059

Vesco Medical Llc

Summary

The FDA issued a Class II for Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059 by Vesco Medical Llc. Reason: The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-.

Details

Source

Device Recall

External ID

Z-0379-2027

Action Date

2026-11-18

Status

Ongoing

Category

device

Product Description

Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059

Lot/Code Info: UDI-DI 00810096151377; Lot Numbers: all lots manufactured through 03/26/2026

Quantity Affected: 2,220 units

Reason for Recall

The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-10-08

Company

Vesco Medical Llc

Westerville, OH

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 188 device recalls issued in the same week, part of 475 device-related FDA actions this month.

Vesco Medical Llc has 8 FDA actions in our database, including 7 recalls and 1 clearance.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Vesco Medical Llc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Vesco Medical Llc have FDA actions?

Vesco Medical Llc has 8 FDA actions in our database, including 7 recalls and 1 device clearance.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0379-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.