RecallHawk
Class II Recall

Atellica CH HDL Cholesterol (HDLC); Siemens Material Number (SMN): 11537213;

Siemens Healthcare Diagnostics, Inc.

Summary

The FDA issued a Class II for Atellica CH HDL Cholesterol (HDLC); Siemens Material Number (SMN): 11537213; by Siemens Healthcare Diagnostics, Inc.. Reason: Reagent probe carryover on the Atellica CH and Atellica CI Analyzers may cause falsely elevated Hemoglobin A1c (A1c_E/A1c_H) results due to HDLC carry.

Details

Source

Device Recall

External ID

Z-0364-2027

Action Date

2026-11-18

Status

Ongoing

Category

device

Product Description

Atellica CH HDL Cholesterol (HDLC); Siemens Material Number (SMN): 11537213;

Lot/Code Info: Siemens Material Number (SMN): 11537213; UDI-DI: 00630414610832; All lots;

Quantity Affected: 72207 units (9744 US, 62463 OUS)

Reason for Recall

Reagent probe carryover on the Atellica CH and Atellica CI Analyzers may cause falsely elevated Hemoglobin A1c (A1c_E/A1c_H) results due to HDLC carryover (biases of 9 29% in NGSP units), and falsely depressed Direct Bilirubin 2 (DBil_2) results due to TP_2 carryover (biases of 0% to -25%), which may affect patient samples, quality control, and/or calibrator results.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cura¿ao, St Eus, Czech Republic, Democratic Republic Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Mexico, Morocco, Nepal, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E., Uganda, United Kingdom , Uruguay, Uzbekistan, Vatikancity, Vietnam.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-09-22

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 188 device recalls issued in the same week, part of 475 device-related FDA actions this month.

Siemens Healthcare Diagnostics, Inc. has 159 FDA actions in our database, including 137 recalls and 22 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Siemens Healthcare Diagnostics, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Siemens Healthcare Diagnostics, Inc. have FDA actions?

Siemens Healthcare Diagnostics, Inc. has 159 FDA actions in our database, including 137 recalls and 22 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0364-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions