Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes:
Summary
The FDA issued a Class II for Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979 by Cochlear Limited. Reason: Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits.
Details
Source
Device Recall
External ID
Z-0349-2027
Action Date
2026-11-11
Status
Ongoing
Category
device
Product Description
Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.
Lot/Code Info: Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286
Quantity Affected: 4 device in the U.S.
Reason for Recall
Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
Distribution
U.S.: NJ, NY, and PA. O.U.S.: Not provided
Type: Voluntary: Firm initiated
Recall Initiated: 2026-09-30
Company
Macquarie University
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 127 device recalls issued in the same week, part of 315 device-related FDA actions this month.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Cochlear Limited) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Cochlear Limited have FDA actions?
This is the only FDA action we have on record for Cochlear Limited in our database.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-0349-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;